Improved Mass Determination of Poly(ethylene glycols) by Electrospray Ion-Mobility Time-of-Flight Mass Spectrometry Coupled with Ion–Molecule Reactions
The authors developed a method to accurately measure the average molecular weight of large poly(ethylene glycols) (PEGs) using ion-mobility time-of-flight mass spectrometry coupled with gas-phase ion–molecule reactions.


Pharmaceutical Technology


Experiment

Chemicals and reagents. Methanol (CH3OH) and isopropanol (IPA) were purchased from Fisher Scientific (Pittsburgh, PA). Ultrapure water (18.2 M) was obtained from a Milli-Q water purifier (Millipore Corporation, Medford, MA). Ammonium acetate, cesium iodide and superbase, 1,8-diazabicyclo[5.4.0]-undec-7-ene (DBU), were purchased from Sigma (St. Louis, MO). PEG 4450 was obtained from Waters Corporation (Milford, MA). PEG-aldehyde 20K was purchased from NOF Corporation (Tokyo, Japan). All materials were used without additional purification.

Sample preparation. All polymer samples were prepared at analyte concentrations of 1–2 mg/mL (10-4 M based on the nominal molecular mass) either in a 50/50 (v/v) solution of water and methanol or 10 mM ammonium acetate and methanol. The solutions were directly infused into the ESI source at 10 µL/min by a syringe pump (Harvard Apparatus, Holliston, MA).

Mass spectrometry. All experiments were performed on a Synapt high-definition mass spectrometer, a hybrid quadrupole ion-mobility TOF mass spectrometer (Waters Corp.). Detailed descriptions of the instrument was found elsewhere (17, 18). A schematic diagram of the instrument is shown in Figure 1a.


Figure 1a & 1b
The instrument was operated in positive-ion mode with a capillary voltage of 3.0 kV. The sampling cone voltage was set at 40 V. The ion source block and nitrogen desolvation gas temperatures were set to 100 °C and 250 °C, respectively. Nitrogen was used as both the nebulizing gas and the desolvation gas. The desolvation gas was set to a flow rate of 250 L/h. The quadrupole was operated in nonresolving mode to transmit a wide m/z range.

The instrument can be operated in two modes: TOF mode and TOF-mobility mode. For TOF-mobility experiments, the ion mobility cell pressure was maintained at 0.5 mbar (N2) and separations were optimized by tuning the wave pulse height from 5 to 14 V. The transfer cell pressure was maintained at 0.02 mbar unless stated otherwise. Pressures in the trap and IMS cell were the direct readbacks from the active pirani gauges equipped with the instrument. The RF amplitude used on the trap, ion mobility cell, and transfer cell was 320 V peak-to-peak. Ion mobility cell traveling wave velocities of 240 m/s were used.

The instrument was externally mass calibrated with cesium iodide (1.0 mg/mL in 50% IPA–50% water). Data acquisition and processing were controlled by MassLynx 4.1 software (Waters Corp). Mass spectra were acquired every 1 s in TOF mode and every 2 s in TOF-mobility mode from m/z 100 to m/z 6000 for PEG 4450 samples and from m/z 500 to m/z 10,000 for PEG-aldehyde 20K for 6 min. Additional data processing to yield the information on the average molecular weight and the polydispersity of the PEGs followed the methods described below.


ADVERTISEMENT

post a comment
Your email address will NOT be published.
appears with your comment
read our privacy policy
Note: does not support HTML
All comments submitted are subject to review, and may be delayed before posting. We reserve the right not to post comments.
LCGC E-mail Newsletters
Subscribe:Check the box next to each newsletter you wish to subscribe for:
| Weekly
| Monthly
| Monthly | Weekly
 Name:
 E-mail:

Survey
At this month's AAPS Annual Meeting & Exposition, several graduate students were awarded as young innovators. What are your thoughts on innovation and the next generation?
We need more young innovators and more students enrolling in pharma science programs
Our industry has just the right mix of young and old to keep innovation going
Age is irrelevant when it comes to innovation
We need more young innovators and more students enrolling in pharma science programs
25%
Our industry has just the right mix of young and old to keep innovation going
16%
Age is irrelevant when it comes to innovation
59%
View Results
Jim MillerOutsourcing OutlookJim Miller Facing Reality
Patricia Van ArnumIngredients InsiderPatricia Van Arnum Advancing Chiral Chemistry in API Synthesis
Faiz kermaniSpotlightFaiz Kermani Demographic time bomb
Rx-360 Takes on Europe, Talks to PharmTech in Podcast Series
Sanctions, Not Barcodes
Vaccines Finding Their Way in Novel Applications
Genzyme’s Friday the 13th
DTCA: Beneficial or Harmful?
FindPharma Custom Search
Source: Pharmaceutical Technology,
Click here