Comparing Calibration Technologies for Liquid-Handling Quality Assurance
The authors consider several common techniques for verifying the accuracy of liquid-handling equipment and offer guidance for finding the appropriate technique for a given instrument.


Pharmaceutical Technology
Volume 32, Issue 8

Ratiometric photometry

The ratiometric photometric calibration method is a refinement of photometry designed to overcome the accuracy limitations of traditional single-dye photometric volume measurements. Ratiometric photometry uses two standardized dyes, and its measurement process produces absorbance readings in pairs that can be combined into absorbance-ratio readings.

The primary benefit of this approach is its ability to improve the accuracy and robustness of measurement comparison with nonratiometric methods. Absorbance ratios can be measured more accurately than individual absorbances, leading to a higher degree of accuracy and precision in ratiometric methods versus traditional single-dye photometric methods. The underlying reason for the improved measurement ability of ratiometric photometric methods is that the absorbance of photometric calibration standards drifts over time, while ratios exhibit greater stability.

Compared to gravimetry, this method offers greater speed, ease-of-use, and enhanced accuracy in small-volume measurements. Compared to fluorometry, ratiometric photometry provides accuracy as well as precision measurements and can do so to a traceable standard because the dyes function as an internal standard. Measuring the second dye in comparison to the first dye provides a nearly automatic compensation for the most common photometric error sources.

Systems based on ratiometric photometry provide information about each individual channel in multichannel devices and good plate-to-plate reproducibility. For ratiometric photometry to produce benefits, however, it must use well-characterized plates and carefully calibrated solutions of good stability.

In addition, to function properly, ratiometric photometric methods require specially formulated dyes to produce accurate absorbance ratios. Lastly, this technology is not always preferred when measuring only large volumes because other technologies may produce adequate measurements more cost effectively.

In summary, ratiometric photometry calibrations provide great benefits when measuring small liquid volumes for protocols requiring traceability and a high degree of accuracy per channel as well as precision.

Summary

Pharmaceutical laboratories have varying protocols, processes, and requirements, and these elements can affect the choice of calibration technologies for liquid-handling devices. Gravimetry, fluorometry, single-dye photometry, and ratiometric photometry are common means for verifying liquid-handling instrumentation. Each technique has its own advantages and disadvantages. Understanding the assay and laboratory quality requirements, traceability needs, and tolerance for error as well as the level of accuracy and precision required can help laboratories make the right decision.

Richard Curtis, PhD*, is chairman and chief technology officer, and George Rodrigues, PhD, is a senior scientific manager at ARTEL, 25 Bradley, Westbrook, ME 04092, tel. 207.854.0860, fax 207.854.0867,
.

*To whom all correspondence should be addressed.

Submitted: Jan. 14, 2008. Accepted: Feb. 4, 2008.

References

1. ASTM International, "ASTM E1154-89 (2003) Standard Specification for Piston or Plunger Operated Volumetric Apparatus," (West Conshohocken, PA, 2003).

2. International Organization for Standardization, "ISO 8655-6: Piston-Operated Volumetric Apparatus—Part 6: Gravimetric Methods for the Determination of Measurement Error" (Geneva, 2002).

3. P. Taylor et al., "A Standard Operating Procedure for Assessing Liquid Handler Performance in High-Throughput Screening," J. Biomol. Screen. 7 (6), 554–569, (2002).

4. J. Petersen and J. Nguyen, "Comparison of Absorbance and Fluorescence Methods for Determining Liquid Dispensing Precision," JALA 10 (2), 82–87, (2005).

5. ISO, "ISO 8655-7: Piston-Operated Volumetric Apparatus—Part 7: Non-Gravimetric Methods for the Assessment of Equipment Performance" (Geneva, 2005).




What would you do differently? Email your thoughts about this paper to
and we may post them to the site.


ADVERTISEMENT

post a comment
Your email address will NOT be published.
appears with your comment
read our privacy policy
Note: does not support HTML
All comments submitted are subject to review, and may be delayed before posting. We reserve the right not to post comments.
LCGC E-mail Newsletters
Subscribe:Check the box next to each newsletter you wish to subscribe for:
| Weekly
| Monthly
| Monthly | Weekly
 Name:
 E-mail:

Survey
At this month's AAPS Annual Meeting & Exposition, several graduate students were awarded as young innovators. What are your thoughts on innovation and the next generation?
We need more young innovators and more students enrolling in pharma science programs
Our industry has just the right mix of young and old to keep innovation going
Age is irrelevant when it comes to innovation
We need more young innovators and more students enrolling in pharma science programs
25%
Our industry has just the right mix of young and old to keep innovation going
16%
Age is irrelevant when it comes to innovation
59%
View Results
Jim MillerOutsourcing OutlookJim Miller Facing Reality
Patricia Van ArnumIngredients InsiderPatricia Van Arnum Advancing Chiral Chemistry in API Synthesis
Faiz kermaniSpotlightFaiz Kermani Demographic time bomb
Rx-360 Takes on Europe, Talks to PharmTech in Podcast Series
Sanctions, Not Barcodes
Vaccines Finding Their Way in Novel Applications
Genzyme’s Friday the 13th
DTCA: Beneficial or Harmful?
FindPharma Custom Search
Source: Pharmaceutical Technology,
Click here