Dielectric Spectroscopy: Choosing the Right Approach
This tutorial paper is meant to aid in dielectric-sensor selection


Pharmaceutical Technology
Volume 9, Issue 32, pp. 8293


Table I: Results of the ranking algorithm showing the top three frequencies ranked by mean error and standard deviation. Error values are normalized.
in which T is the thickness of the coating, Cf1 is the capacitance measured at frequency f1, and K1, K2, and K3, respectively, are the quadratic, linear, and offset coefficients.

To determine these coefficients and the frequency to be used, a simple algorithm is performed. First, the measured capacitances are averaged over the four samples for each of the coating levels. For each frequency, the minimum capacitance is subtracted to remove offset errors. Then the values are divided by the maximum capacitance at each frequency to remove the linear errors. The resulting capacitance ratios at each frequency are individually fitted to the quadratic equation above, and the resulting error in estimation is computed. The mean of the estimation errors and the standard deviation is tabulated for each frequency. The frequencies are ranked on the basis of least mean error and the least standard deviation. The product of these rankings is used as a figure of merit for selection of the frequency f1. Table I shows the results obtained from the algorithm detailed above. It shows that the measurements at 100 kHz are the most suitable for the estimation process. The coefficients K1, K2, and K3 were determined by least square methods. The resulting equation for estimation of coating thickness is





Figure 6 (ALL FIGURES ARE COURTESY OF THE AUTHORS.)
in which C100k is the mean compensated and normalized capacitance at 100 kHz. Figure 6 shows the comparison between the actual coating thickness and estimated coating thickness. This process of estimation must be further validated by conducting experiments with many more coating levels and samples. Therefore, for inline process measurements, a selected frequency of excitation can be ascertained with some postprocessing of off-line data.


ADVERTISEMENT

post a comment
Your email address will NOT be published.
appears with your comment
read our privacy policy
Note: does not support HTML
All comments submitted are subject to review, and may be delayed before posting. We reserve the right not to post comments.
LCGC E-mail Newsletters
Subscribe:Check the box next to each newsletter you wish to subscribe for:
| Weekly
| Monthly
| Monthly | Weekly
 Name:
 E-mail:

Survey
At this month's AAPS Annual Meeting & Exposition, several graduate students were awarded as young innovators. What are your thoughts on innovation and the next generation?
We need more young innovators and more students enrolling in pharma science programs
Our industry has just the right mix of young and old to keep innovation going
Age is irrelevant when it comes to innovation
We need more young innovators and more students enrolling in pharma science programs
25%
Our industry has just the right mix of young and old to keep innovation going
16%
Age is irrelevant when it comes to innovation
59%
View Results
Jim MillerOutsourcing OutlookJim Miller Facing Reality
Patricia Van ArnumIngredients InsiderPatricia Van Arnum Advancing Chiral Chemistry in API Synthesis
Faiz kermaniSpotlightFaiz Kermani Demographic time bomb
Rx-360 Takes on Europe, Talks to PharmTech in Podcast Series
Sanctions, Not Barcodes
Vaccines Finding Their Way in Novel Applications
Genzyme’s Friday the 13th
DTCA: Beneficial or Harmful?
FindPharma Custom Search
Source: Pharmaceutical Technology,
Click here